1. Device Description & Intended Use
The Medisafe International Reusable Anesthesia Silicone Face Mask is a non-invasive patient interface designed to conduct oxygen, anesthesia gases, or air mixtures directly from a breathing circuit or Manual Silicone Resuscitator (BVM) to the patient’s respiratory tract.
Manufactured from 100% Medical-Grade Liquid Silicone Rubber (LSR), this non-inflatable soft-cushion mask is biocompatible, Latex-Free, and Phthalate-Free. It is intended for multi-patient reuse following validated high-temperature steam sterilization (autoclaving).
2. Indications & Demographic Sizing
The mask line is indicated for positive pressure ventilation across neonatal, pediatric, and adult patient populations during resuscitation and general anesthesia.
|
Size Category |
Target Demographic |
Cushion Shape |
Fitting Standard |
|
Sizes 00, 0, 1 |
Premature / Neonatal / Infant |
Round Dome |
15 mm Female |
|
Size 2 |
Small Pediatric / Child |
Round Dome |
Universal 15/22 mm |
|
Sizes 3, 4, 5 |
Adult (Small / Medium / Large) |
Anatomical Form |
22 mm Female |
3. Warnings, Precautions & Contraindications
- Trained Personnel Only: For use exclusively by qualified healthcare professionals trained in airway management and positive pressure ventilation.
- Ocular Safety Warning: Ensure the mask frame does not rest on or apply downward pressure to the eyes. Ocular compression can induce corneal abrasions or trigger the oculocardiac reflex (bradycardia).
- Aspiration Vigilance: Continually observe patient facial status through the transparent dome. If vomiting or regurgitation occurs, remove the mask immediately and clear the airway.
- Fitting Verification: Verify tight fitment between the mask connection hub and the patient valve before opening oxygen supply or delivering tidal volumes.
4. Reprocessing & Sterilization Protocols
Because this device is multi-patient reusable, it must undergo full cleaning and high-temperature steam sterilization between every patient application.
A. Manual Cleaning Procedure
- Disassembly: Remove the optional colored rigid hook ring from the mask neck (Sizes 3–5).
- Soak & Wash: Submerge mask body and hook ring in warm water mixed with a neutral-pH enzymatic medical detergent. Clean all surfaces with a soft-bristled brush.
- Rinse: Thoroughly rinse components under running purified water to flush all detergent residue.
- Drying: Dry completely with a lint-free cloth or clean compressed air.
B. Steam Autoclave Sterilization (Validated Process)
|
Parameter |
Recommended Autoclave Standard |
Alternative Cycle |
|
Sterilization Type |
Pre-vacuum Steam Autoclave |
Gravity Displacement |
|
Temperature |
134°C |
121°C |
|
Exposure Time |
4 to 5 Minutes |
15 to 20 Minutes |
5. Inspection, Service Life & Storage
- Service Life Limit: Reusable silicone masks are validated to withstand 20 to 30 steam autoclave cycles under standard operational parameters.
- Pre-Use Inspection Criteria: Inspect before every clinical use. Discard and replace immediately if:
- The silicone cushion shows yellowing, clouding, or surface stickiness/tackiness.
- Cracks, tearing, or physical warping are visible on the sealing edge.
- Storage: Store reprocessed masks in clean, dry sealed packages away from direct sunlight, chemical fumes, and extreme temperature fluctuations.