IFU ! Reusable Anesthesia Silicone Face Masks

  • Home
  • IFU ! Reusable Anesthesia Silicone Face Masks

IFU ! Reusable Anesthesia Silicone Face Masks

Instructions for Use (IFU) ! Reusable Anesthesia Silicone Face Masks

1. Device Description & Intended Use

The Medisafe International Reusable Anesthesia Silicone Face Mask is a non-invasive patient interface designed to conduct oxygen, anesthesia gases, or air mixtures directly from a breathing circuit or Manual Silicone Resuscitator (BVM) to the patient’s respiratory tract.

Manufactured from 100% Medical-Grade Liquid Silicone Rubber (LSR), this non-inflatable soft-cushion mask is biocompatible, Latex-Free, and Phthalate-Free. It is intended for multi-patient reuse following validated high-temperature steam sterilization (autoclaving).

2. Indications & Demographic Sizing

The mask line is indicated for positive pressure ventilation across neonatal, pediatric, and adult patient populations during resuscitation and general anesthesia.

Size Category

Target Demographic

Cushion Shape

Fitting Standard

Sizes 00, 0, 1

Premature / Neonatal / Infant

Round Dome

15 mm Female

Size 2

Small Pediatric / Child

Round Dome

Universal 15/22 mm

Sizes 3, 4, 5

Adult (Small / Medium / Large)

Anatomical Form

22 mm Female

3. Warnings, Precautions & Contraindications

  • Trained Personnel Only: For use exclusively by qualified healthcare professionals trained in airway management and positive pressure ventilation.
  • Ocular Safety Warning: Ensure the mask frame does not rest on or apply downward pressure to the eyes. Ocular compression can induce corneal abrasions or trigger the oculocardiac reflex (bradycardia).
  • Aspiration Vigilance: Continually observe patient facial status through the transparent dome. If vomiting or regurgitation occurs, remove the mask immediately and clear the airway.
  • Fitting Verification: Verify tight fitment between the mask connection hub and the patient valve before opening oxygen supply or delivering tidal volumes.

4. Reprocessing & Sterilization Protocols

Because this device is multi-patient reusable, it must undergo full cleaning and high-temperature steam sterilization between every patient application.

A. Manual Cleaning Procedure

  1. Disassembly: Remove the optional colored rigid hook ring from the mask neck (Sizes 3–5).
  2. Soak & Wash: Submerge mask body and hook ring in warm water mixed with a neutral-pH enzymatic medical detergent. Clean all surfaces with a soft-bristled brush.
  3. Rinse: Thoroughly rinse components under running purified water to flush all detergent residue.
  4. Drying: Dry completely with a lint-free cloth or clean compressed air.

B. Steam Autoclave Sterilization (Validated Process)

Parameter

Recommended Autoclave Standard

Alternative Cycle

Sterilization Type

Pre-vacuum Steam Autoclave

Gravity Displacement

Temperature

134°C

121°C

Exposure Time

4 to 5 Minutes

15 to 20 Minutes

5. Inspection, Service Life & Storage

  • Service Life Limit: Reusable silicone masks are validated to withstand 20 to 30 steam autoclave cycles under standard operational parameters.
  • Pre-Use Inspection Criteria: Inspect before every clinical use. Discard and replace immediately if:
    1. The silicone cushion shows yellowing, clouding, or surface stickiness/tackiness.
    2. Cracks, tearing, or physical warping are visible on the sealing edge.
  • Storage: Store reprocessed masks in clean, dry sealed packages away from direct sunlight, chemical fumes, and extreme temperature fluctuations.